Steffen Brinckmann authoredSteffen Brinckmann authored
Code owners
Assign users and groups as approvers for specific file changes. Learn more.
Doli_Tensile_Test.md 1.29 KiB
Tensile testing with Doli
- Setup controll box at instrument
- General information
- buttons F1..F3 correspond to three symbols above
- PC-mode: use for control by PC
- turn knob to get there and then press ESC to get move the control from the handheld to the PC
- Start Software and hope that LAN connection works
- otherwise retry
- Button Bike "Travel order"
- distance-controlled, relative: 100mm with 100mm/min
- direction clear
- Measure the sample:
- |width||
- |thickness||
- Mount sample: top first and then move into bottom
- Button Setup experiment: Hammer button
- "Test procedure, speeds", "End of test" most important
- Switch from F0->F1 confusing
- choose F0 as max force
- Example speeds:
- metals=0.5mm/min
- polymers=5mm/min
- Adopt path: "Store data to"
- all 3 tare-buttons should be pressed
- Button Traffic-Light: start experiments
- Button Diagram (one line)
- open & close: rescale graph
- After test: Travel order
- go back relative to 0
- After all is done
- Close all wedges softly
- drive back to touch almost: s=35mm
- Plot: click on series to plot all
- double click on test
- export by right click in graph and then copy-to-txt
- Results in C:\Versuchsergebnisse
Author: Steffen Brinckmann, IMD-1, FZJ