Skip to content
Snippets Groups Projects
Commit 71a4762b authored by Timo Bisswanger's avatar Timo Bisswanger
Browse files

Update Stanford_SR560.ini

Adapted and extended. Added listen all etc. and default settings.
Changed termination to LF as it is obviously necessary for Labber ver. 1.7.7
parent 7c5d4b37
No related branches found
No related tags found
1 merge request!12Ag spin general
# Instrument driver configuration file.
#2019-01-31, Timo set up from SRS830
#
#This device is listen only!
#The interface is configured as listen-only,
#########9600 baud DCE, 8-bit, no parity, 2 stop bits, ############
#and is optically isolated to prevent any noise or grounding problems.
#
# Manual says: Commands must end with a carriage return and line feed <CR><LF>.
# But currently with the Labber Ver. 1.7.7 you must configure the serial communication with <LF> only!!!!
#use USB-to-RS232 and set the before mentioned parameters also in the COM driver of the operating system, Flow Controle=none
[General settings]
......@@ -9,7 +17,7 @@
name: Stanford Low-Noise Preamplifier SR 560
# The version string should be updated whenever changes are made to this config file
version: 1.0
version: 1.1
# Name of folder containing the code defining a custom driver. Do not define this item
# or leave it blank for any standard driver based on the built-in VISA interface.
......@@ -43,7 +51,8 @@ reset: True
timeout: 5
# Default termination character
term_char: CR+LF
#in Labber ver. 1.7.7 just LF, from Manual it should be CR+LF
term_char: LF
# Query instrument errors (True or False). If True, every command sent to the device will
# be followed by an error query. This is useful when testing new setups, but may degrade
......@@ -67,7 +76,7 @@ error_cmd:
# Initialization commands are sent to the instrument when starting the driver
# *RST will reset the device, *CLS clears the interface
init:
init:LALL
# Final commands sent to the instrument when closing the driver
final:
......@@ -144,7 +153,7 @@ group: Input
section: Input and Filter
permission: WRITE
[Filter mode] #should be fine
[Filter mode]
datatype: COMBO
def_value: 12 dB low pass
combo_def_1: bypass
......@@ -164,7 +173,7 @@ group: Filters
section: Input and Filter
permission: WRITE
[Highpass Cutoff] #fine?
[Highpass Cutoff]
#HFRQi Sets highpass filter frequency.
#I = 0 – 11 sets frequency = 0.03
#Hz to 10 kHz
......@@ -200,7 +209,7 @@ section: Input and Filter
show_in_measurement_dlg: True
permission: WRITE
[Lowpass Cutoff] #fine?
[Lowpass Cutoff]
#LFRQ i Sets lowpass filter frequency.
#i = 0 – 15 sets frequency
#= 0.03 Hz to 1 MHz
......@@ -306,8 +315,8 @@ permission: WRITE
datatype: BUTTON
set_cmd: *RST
#Reset. Recalls default settings.
group: Blank and reset
section: others
group: Reset
section: Reset and Listen
permission: WRITE
tooltip: Recalls default settings.
......@@ -315,29 +324,64 @@ tooltip: Recalls default settings.
datatype: BUTTON
set_cmd: ROLD
#ROLD Resets overload for ½ second.
group: Blank and reset
section: others
group: Reset
section: Reset and Listen
permission: WRITE
tooltip: Resets overload for ½ second.
#[Listen All]
#LALL Listen all. Makes all attached
#SR560’s listeners.
#group: Listen
#section: others
[Listen All]
datatype: BUTTON
set_cmd: LALL
#Reset. Recalls default settings.
group: Listen
section: Reset and Listen
permission: WRITE
tooltip: Sets all connected SR560 to listen.
#[Make listener]
#LISN i Listen command. Makes
#SR560 with address i (0,1,2,3)
#a listener.
#group: Listen
#section: others
[Unlisten All]
datatype: BUTTON
set_cmd: UNLS
#Reset. Recalls default settings.
group: Listen
section: Reset and Listen
permission: WRITE
tooltip: Set unlisten to all connected SR560.
#[Unlisten]
#UNLS Unlisten. Unaddresses all
#attached SR560’s.
#group: Listen
#section: others
[Listen]
datatype: COMBO
def_value: 0
combo_def_1: 0
combo_def_2: 1
combo_def_3: 2
combo_def_4: 3
cmd_def_1: 0
cmd_def_2: 1
cmd_def_3: 2
cmd_def_4: 3
set_cmd: LISN
group: Listen
section: Reset and Listen
permission: WRITE
tooltip: Sets selected SR560 to listen.
[SV Defaults]
datatype: BUTTON
set_cmd: LALL;CPLG 0;GAIN 9;FLTM 2;LFRQ 12;SRCE 1;DYNR 0;
#Reset. Recalls default settings.
group: SV and Dirac
section: Defaults
permission: WRITE
tooltip: Set default values for SV measurements.
[Set Defaults]
datatype: BUTTON
set_cmd: LALL;CPLG 0;GAIN 0;FLTM 2;LFRQ 12;SRCE 1;DYNR 0;
#Reset. Recalls default settings.
group: SV and Dirac
section: Defaults
permission: WRITE
tooltip: Set default values for Dirac measurements.
#[Vernier Gain Status]
#UCAL i Sets the vernier gain status.
......